CASE REPORT |
https://doi.org/10.5005/jp-journals-10071-24219 |
Cardiac Pacing a Rare Cause of Ventilator Auto Triggering
1,2Department of Critical Care Medicine, Narayana Hrudayalaya, Bengaluru, Karnataka, India
Corresponding Author: Nithish Mukunthan M, Department of Critical Care Medicine, Narayana Hrudayalaya, Bengaluru, Karnataka, India, Phone: +91 9789666562, e-mail: nithishchat@gmail.com
How to cite this article: Mukunthan MN, Bhardwaj V. Cardiac Pacing a Rare Cause of Ventilator Auto Triggering. Indian J Crit Care Med 2022;26(5):643–645.
Source of support: Nil
Conflict of interest: None
ABSTRACT
Ventilator auto triggering is an avoidable complication in ventilators, if left unnoticed can lead to deleterious effects. There are various causes for ventilator auto triggering. Though rare, there are some cardiac causes for inadvertent ventilator triggering. We report a case of 44-years-old male paced with atrial epicardial wires postcoronary artery bypass. The wires were close to the right phrenic nerve, causing the right diaphragm to contract in synchronization with the heartbeat. This caused ventilator auto triggering and ended up delivering inadvertent breaths. The pacemaker output was immediately reduced to the required minimum to continue pacing the heart and decrease phrenic nerve stimulation. This caused immediate changes in ventilator waveform and auto triggering was completely stopped and the patient could be successfully weaned off the ventilator. This case report emphasizes he need for timely recognition of alteration in ventilator waveforms and early intervention to avert any untoward events.
Keywords: Epicardial pacing, Flow trigger, Inadvertent triggering, Pacemaker, Respiratory alkalosis, Ventilator auto triggering, Ventilator waveforms.
HIGHLIGHTS
Cardiac pacing is a rare but noteworthy cause of ventilator auto triggering.
Happens when pacing wires are in close proximation to the right phrenic nerve.
Timely recognition of auto trigger waveforms and early necessary interventions can avert untoward events.
INTRODUCTION
Mechanical ventilators may sometime deliver additional breathes to a patient due to unintentional triggering which is called ventilator auto triggering.1 There are various reasons for auto triggering like water condensation in the circuit, stuck valves, leak in any portion of circuit or cuff leaks, decreased trigger settings in ventilator, inadvertent patient triggering, etc.2–4 Though rare, there are some cardiac causes for inadvertent ventilator triggering like cardiac oscillations and external pacemakers.5–7 Among pacemakers, the most prevalent cause is when atria are paced. This ventilator auto triggering causes multiple complications in patients like respiratory alkalosis, auto peeping, delayed ventilation, increased work of breathing, etc.8,9 We report a case of ventilator auto triggering in a case of postoperative coronary artery bypass who was put on atrial pacing and its management. It also gives us an insight into the need for timely recognition of alteration in ventilator graphs, troubleshooting the causes for variations in graphs, and early interventions to avert any untoward events.
CASE
A 44-year male known case of diabetes and chronic kidney disease was diagnosed with coronary artery disease and underwent a left anterior descending artery (LAD) endarterectomy plus coronary artery bypass with two grafts and two atrial and ventricular epicardial wires were placed (Fig. 1). Postoperatively the patient was received in the intensive care unit (ICU) intubated with minimal pressure. He was put on Mindray Synovent ES-mechanical ventilator on pressure regulated volume control mode with fractional inspiratory oxygen concentration (FiO2)—40%, tidal volume—450 mL, flow trigger—2.0, positive end expiratory pressure (PEEP)—5, respiratory rate—14/minute. He was deeply sedated and ventilated overnight.
Fig. 1: Intraoperative placement of two atrial and ventricular epicardial wires
In the night he was started on atrial pacing for sinus bradycardia with a temporary cardiac pacemaker with sensitivity—6 millivolt (mV), output—12 volt (V), rate—90/minute. The next day morning sedation was decreased and while the patient was planning for extubation, we noticed auto triggering of the ventilator. The flow-time and pressure-time scalar showed multiple triggers in both phases of respiration which was triggering inadvertent breaths. On further examination of the patient there was a spontaneous contraction of the diaphragm with each heartbeat (Video 1). We hypothesized that the current from the atrial pacing wires could be causing right phrenic nerve stimulation which runs close by causing this entire phenomenon. We decreased the pacemaker output from 12 to 4 V (Fig. 2) which immediately stopped the diaphragm contraction and the auto triggering thereby normalizing the ventilator graphs (Fig. 3) (Video 2).
Fig. 2: Reducing the pacemaker output from 12 to 4 V
Fig. 3: The waveform is in pressure-regulated volume control mode. The red arrow shows the auto triggering waveforms with 12 V pacemaker output and the yellow arrow shows the resolution of waveforms once output was reduced to 4 V
For confirmation we repeated the same maneuver with pressure support ventilation with flow trigger—2.0, pressure support—12, FiO2—40%, PEEP—5 and could elicit the same results (Fig. 4). Finally, the pacemaker output was reduced to 4 V and the patient was put on pressure support ventilation and extubated at the earliest avoiding any untoward complications.
Fig. 4: Waveform in pressure support ventilation mode. The red arrow shows the auto triggering waveforms with 12 V pacemaker output and the yellow arrow shows the resolution of waveforms once output was reduced to 4 V
DISCUSSION
This case report demonstrates the importance of timely recognition of ventilator auto triggering and rectifying the cause for it to prevent untoward complications and delayed weaning. Artificial pacemakers are used to depolarize the atria and/or ventricle and make them contract. Usually, in post-cardiac surgery, atrial pacing is the most common form of pacing.7 In many cases the epicardial atrial pacing leads are kept close to the right phrenic nerve intraoperatively for better sensing.10 And postoperatively for better pacing, it is a common practice to keep the output higher as there may be surrounding tissue injury around the pacing leads resulting in variable energy output.11 All these end up stimulating the right phrenic nerve and causing right diaphragm contraction and causing ventilator auto triggering. Such auto triggering can result in many complications like hyperventilation, respiratory alkalosis, auto peeping, hypotension, decreased urine output, delayed weaning, increased work of breathing, etc.8,9
Bedside ventilator waveform analysis is an important skill for intensivists. By timely interpretation of graphs and decreasing the pacemaker output to the minimum necessary to maintain cardiac contraction, all the complications of inadvertent ventilator auto triggering were successfully averted.
CONCLUSION
Cardiac pacing as a cause of ventilator auto triggering is a very rare case scenario but clinically significant as failure to diagnose may cause complications that could have been avoided otherwise. This case report also emphasizes the importance of bedside recognition of various ventilator waveforms for early intervention and good patient outcomes.
SUPPLEMENTARY MATERIAL
All the supplementary videos are available on www.ijccm.org.
Video 1: Spontaneous diaphragmatic contraction corresponding to each heartbeat triggered by the pacemaker with 12 V output
Video 2: Waveforms in pressure regulated volume control mode showing auto triggering with 12 V pacemaker output which resolves to regular waveforms once the output is reduced to 4 V
ORCID
Nithish Mukunthan M https://orcid.org/0000-0003-4252-2766
Vimal Bhardwaj https://orcid.org/0000-0003-4273-2771
REFERENCES
1. France TE, Garnett AR, Grooms D. Inadvertent triggering of the ventilator caused by surgically placed pacer wires. Respir Care 2012;57(8):1334–1336. DOI: 10.4187/respcare.01378.
2. Pachisia AV, Bhattacharyya S. Post-cardiac arrest ventilator triggering. Indian J Crit Care Med 2019;23(4):196. DOI: 10.5005/jp-journals-10071-23155.
3. Sassoon C. Triggering of the ventilator in patient-ventilator interactions. Respir Care 2011;56(1):39–48. DOI: 10.4187/respcare.01006.
4. Schawb RJ, Schnader J. Ventilator autocycling due to an endotracheal tube cuff leak. Chest 1991;100(10):1172–1173. DOI: 10.1378/chest.100.4.1172.
5. Kaloria N, Gupta A, Goila A, Sood R. Cardiogenic oscillation induced ventilator autotriggering. J Neuroanaesthesiol Crit Care 2015;2(02):134–135. DOI: 10.4103/2348-0548.154239.
6. Imanaka H, Nishimuram N, Takeuchi M, Kimball R, Yahagi N, Kumon K. Autotriggering caused by cardiac oscillation during flow-triggered mechanical ventilation. Crit Care Med 2000;28(2):402–407. DOI: 10.1097/00003246-200002000-00019.
7. Khan AA, Nash A, Ring NJ, Marshall AJ. Right hemidiaphragmatic twitching: a complication of bipolar atrial pacing. Pacing Clin Electrophysiol 1997;20(6):1732–1733. DOI: 10.1111/j.1540-8159.1997.tb03550.x.
8. Oakes D. Clinical practitioner’s pocket guide to respiratory care. Old Town, ME: Health Educator Publications; 2000. p. 2–14.
9. Sager JS, Eiger G, Fuchs BD. Ventilator auto-triggering in a patient with tuberculous bronchopleural fistula. Respir Care 2003;48(5):519–521. PMID: 12729469.
10. Wells WJ, Batra AS. Management of phrenic nerve stimulation caused by epicardial pacemaker leads in children. Ann Thorac Surg 2003;76(5):1757–1758. DOI: 10.1016/s0003-4975(03)00557-5.
11. Safavi-Naeini P, Saeed M. Pacemaker troubleshooting: common clinical scenarios. Tex Heart Inst J 2016;43(5):415–418. DOI: 10.14503/THIJ-16-5918.
________________________
© The Author(s). 2022 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by-nc/4.0/), which permits unrestricted use, distribution, and non-commercial reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.