Cardiac Pacing a Rare Cause of Ventilator Auto Triggering
M Nithish Mukunthan, Vimal Bhardwaj
Citation Information :
Mukunthan MN, Bhardwaj V. Cardiac Pacing a Rare Cause of Ventilator Auto Triggering. Indian J Crit Care Med 2022; 26 (5):643-645.
Ventilator auto triggering is an avoidable complication in ventilators, if left unnoticed can lead to deleterious effects. There are various causes for ventilator auto triggering. Though rare, there are some cardiac causes for inadvertent ventilator triggering. We report a case of 44-years-old male paced with atrial epicardial wires postcoronary artery bypass. The wires were close to the right phrenic nerve, causing the right diaphragm to contract in synchronization with the heartbeat. This caused ventilator auto triggering and ended up delivering inadvertent breaths. The pacemaker output was immediately reduced to the required minimum to continue pacing the heart and decrease phrenic nerve stimulation. This caused immediate changes in ventilator waveform and auto triggering was completely stopped and the patient could be successfully weaned off the ventilator. This case report emphasizes he need for timely recognition of alteration in ventilator waveforms and early intervention to avert any untoward events.
France TE, Garnett AR, Grooms D. Inadvertent triggering of the ventilator caused by surgically placed pacer wires. Respir Care 2012;57(8):1334–1336. DOI: 10.4187/respcare.01378.
Pachisia AV, Bhattacharyya S. Post-cardiac arrest ventilator triggering. Indian J Crit Care Med 2019;23(4):196. DOI: 10.5005/jp-journals-10071-23155.
Sassoon C. Triggering of the ventilator in patient-ventilator interactions. Respir Care 2011;56(1):39–48. DOI: 10.4187/respcare.01006.
Schawb RJ, Schnader J. Ventilator autocycling due to an endotracheal tube cuff leak. Chest 1991;100(10):1172–1173. DOI: 10.1378/chest.100.4.1172.
Kaloria N, Gupta A, Goila A, Sood R. Cardiogenic oscillation induced ventilator autotriggering. J Neuroanaesthesiol Crit Care 2015;2(02):134–135. DOI: 10.4103/2348-0548.154239.
Imanaka H, Nishimuram N, Takeuchi M, Kimball R, Yahagi N, Kumon K. Autotriggering caused by cardiac oscillation during flow-triggered mechanical ventilation. Crit Care Med 2000;28(2):402–407. DOI: 10.1097/00003246-200002000-00019.
Khan AA, Nash A, Ring NJ, Marshall AJ. Right hemidiaphragmatic twitching: a complication of bipolar atrial pacing. Pacing Clin Electrophysiol 1997;20(6):1732–1733. DOI: 10.1111/j.1540-8159.1997.tb03550.x.
Oakes D. Clinical practitioner's pocket guide to respiratory care. Old Town, ME: Health Educator Publications; 2000. p. 2–14.
Sager JS, Eiger G, Fuchs BD. Ventilator auto-triggering in a patient with tuberculous bronchopleural fistula. Respir Care 2003;48(5):519–521. PMID: 12729469.
Wells WJ, Batra AS. Management of phrenic nerve stimulation caused by epicardial pacemaker leads in children. Ann Thorac Surg 2003;76(5):1757–1758. DOI: 10.1016/s0003-4975(03)00557-5.
Safavi-Naeini P, Saeed M. Pacemaker troubleshooting: common clinical scenarios. Tex Heart Inst J 2016;43(5):415–418. DOI: 10.14503/THIJ-16-5918.