VOLUME 17 , ISSUE 2 ( April, 2013 ) > List of Articles

LETTER TO EDITOR

Comparison of central venous saturation by standard ABG machine versus co-oximeter: Is 18 carat as good as the 24 carat gold standard?

Gayathri Subramanian, V Anitha

Keywords : Central venous saturation, co-oximeter, goal directed therapy, septic shock

Citation Information : Subramanian G, Anitha V. Comparison of central venous saturation by standard ABG machine versus co-oximeter: Is 18 carat as good as the 24 carat gold standard?. Indian J Crit Care Med 2013; 17 (2):82-86.

DOI: 10.4103/0972-5229.114824

License: CC BY-ND 3.0

Published Online: 01-06-2014

Copyright Statement:  Copyright © 2013; The Author(s).


Abstract

Aims: Aggressive therapy aimed at desired end-points of Early Goal Directed Therapy (EGDT) is the cornerstone of septic shock management. A key endpoint that improves outcomes is achieving central venous saturation (ScvO 2 ) >70%. The gold standard to measure ScvO 2 is by a co-oximeter (co-ox). Settings and Design: This prospective, observational study from a multidisciplinary pediatric intensive care unit (PICU) was conducted to assess the validity of ScvO 2 levels by standard ABG (stand ABG) machine in comparison with co-ox in conditions that shifted the oxygen dissociation curve (ODC) to the right or left in sick children and controls. Materials and Methods: Data from paired samples was tested for correlation coefficient for pH, paCO 2, paO 2, and ScvO 2. Tests for correlation (Pearson′s coefficient) and agreement (Bland-Altman analysis) were performed on ScvO 2 values obtained in various subgroups. Sensitivity and specificity for ScvO 2 values determined by standard ABG machine versus co-ox were calculated in reference to EGDT endpoints. Results: A total of 141 paired samples were collected from 82 children. Despite a statistically significant difference in the pH and ScvO 2, there was good linear correlation between these parameters. Limits of agreement between ScvO 2 measured by standard ABG machine and co-ox by Bland-Altman gave 2.3% bias with 95% CI of -24.2% to 19.5%. Sensitivity and specificity of standard ABG machine in detecting low ScvO 2 in shock was 84.21% and 93.18% respectively, while it was false positive in 4 samples. Conclusions: The less expensive standard ABG machine showed satisfactory correlation with gold standard co-ox over a range of patient conditions; however, the wide range for agreement was of concern and it performed particularly poorly in anemic patients.


PDF Share
  1. Dellinger RP, Levy MM, Carlet JM, Bion J, Parker MM, Jaeschke R, et al. Surviving Sepsis Campaign: International guidelines for management of severe sepsis and septic shock: 2008. Crit Care Med 2008;36:296-327.
  2. Brierley J, Carcillo JA, Choong K, Cornell T, Decaen A, Deymann A, et al. Clinical practice parameters for hemodynamic support of pediatric and neonatal septic shock: 2007 update from the American College of Critical Care Medicine. Crit Care Med 2009;37:666-88.
  3. Rasanen J, Peltola K, Leijala M. Superior vena cava and mixed venous oxyhemoglobin saturations in children recovering from open heart surgery. J Clin Monit 1992;8:44-9.
  4. Leone M, Blidi S, Antonini F, Meyssignac B, Bordon S, Garcin F, et al. Oxygen tissue saturation is lower in nonsurvivors than in survivors after early resuscitation of septic shock. Anesthesiology 2009;111:366-71.
  5. Thayssen P, Klarholt E. Relation between caval and pulmonary artery oxygen saturation in children. Br Heart J 1980;43:574-8.
  6. Dueck MH, Klimek M, Appenrodt S, Weigand C, Boerner U. Trends but not individual values of central venous oxygen saturation agree with mixed venous oxygen saturation during varying hemodynamic conditions. Anesthesiology 2005;103:249-57.
  7. Severinghaus JW, Astrup P, Murray JF. Blood gas analysis and critical care medicine. Am J Respir Crit Care Med 1998;157:S114-22.
  8. Woolley A, Hickling K. Errors in measuring blood gases in the intensive care unit: Effect of delay in estimation. J Crit Care 2003;18:31-7.
  9. Twomey PJ, Kroll MH. How to use linear regression and correlation in quantitative method comparison studies. Int J Clin Pract 2008;62:529-38.
  10. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-10.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.